MIL-PRF-38793C
3.2.1 Front matter. The front matter shall be in accordance with the requirements of MIL-STD-
38784.
3.2.2 Section I - Identification and description. The identification and description section shall
include the following information.
3.2.2.1 Test instrument identification. Commercial test instruments shall be identified by
manufacturer, model number, and nomenclature. Military test instruments shall be identified by
model number, nomenclature, and principle function, if not evident from the nomenclature.
Component identification shall include the following information as applicable:
a.
Set and component [(7603) and (7B53A)].
b.
Military item and commercial counterpart [(ANUSM205) and (Hewlett-Packard
650A)].
c.
Model variations (ANPSM45, A).
d.
Any other identification specified by the acquiring activity, such as
manufacturer's part number (see 6.2).
e.
The various configurations, such as cabinet and rack-mount models (i.e.,
Tektronix 545 and RM545) or those using Zener reference or standard cells, such
as Fluke 801 H or 801 HG.
3.2.2.2 Calibration description. The test instrument characteristics, performance specifications,
and test methods pertinent to the calibration, shall be presented in tabular form (see FIGURE 1).
The test method description shall clearly indicate how each test instrument performance
specification is verified, and shall refer to alternate methods if utilized in the procedure.
Variations in test instrument characteristics or performance between models shall be identified.
3.2.2.2.1 Characteristics. Test instrument characteristics shall include those characteristics
identified by the Manufacturer as critical characteristics that require periodic calibration to
ensure the test instrument performance specifications are met. The Calibration Requirements
Test List of the applicable Calibration Requirements Document (CRD) should be reviewed for
any service specific test that may be required (see 6.4.11).
3.2.2.2.2 Performance specifications. Performance specifications shall be consistent with the
item manufacturer's specifications, whenever possible. If they are not, the calibration procedure
shall identify any deviations from the manufacturer's specifications.
3.2.2.3 Accessories. All peculiar accessories that require calibration with the test instrument
shall be identified and described (see 6.4.4).
3.2.3 Section II - equipment requirements. This section shall include a list of the calibration
equipment and peculiar accessories required in the calibration process (see FIGURE 2). The list
shall include the generic name, minimum use specifications (characteristics, range, and accuracy
5